← Bridge Instrumentation & Monitoring

Calibration, Commissioning & Railway Bridge Instrumentation

How sensors and instrumentation systems are calibrated (zero, span, linearity, sensitivity, hysteresis, repeatability), the factory and site acceptance tests (FAT and SAT), commissioning, troubleshooting and periodic maintenance, and the instrumentation of railway bridges under RDSO and Indian Railways practice — strain, deflection and vibration measurement under moving trains and preparation of instrumentation schemes and reports.

📑 Contents (7 sections)

Last reviewed 30 Sept 2026 · 8 min read

Calibration

Calibration is the comparison of an instrument's reading with a reference of known, higher accuracy (traceable to a national standard) and the determination of the correction — the calibration factor — needed to convert its output into engineering units. Without calibration, a reading is only a number.

What is checked

Item Meaning and how it is checked
Zero The reading with no input; corrected by the zero offset
Span / sensitivity The output per unit input (mV/V per kN, mV per mm, Hz² per µε); found by applying known inputs
Linearity The deviation of the readings from the straight line through zero and full scale (as % of full scale)
Hysteresis Difference between readings on the loading and unloading paths
Repeatability Agreement of repeated readings at the same input
Resolution and noise Smallest detectable change and the noise floor
Temperature effect Change of zero and span with temperature
Cross-sensitivity Response to inputs in other directions (a transverse acceleration on an accelerometer)

Methods

  • Laboratory calibration — the load cell is loaded in a calibrated testing machine or dead-weight machine; the LVDT is moved with a micrometer; the accelerometer is compared with a reference on a shaker; thermometers are compared in a temperature bath. Points are taken at increasing and decreasing inputs, at 10–20 % steps, and the calibration certificate records the coefficients and the uncertainty. Laboratories are accredited (in India, NABL) and certificates are traceable to national standards.
  • Shunt calibration (strain gauges and bridge sensors) — a precise resistor is switched across one arm of the bridge to simulate a known strain, and the recorded output is compared with the calculated value; the check includes the gauge, cable and amplifier.
  • In-situ (field) calibration — using known loads (a truck of known weight), a comparison with a portable reference instrument, zero checks under no load, and cross-checks between sensors.
Worked ExampleExample — load cell check

A 1000 kN load cell has a rated sensitivity of 2.0 mV/V. With an excitation of 10 V the full-scale output is 20 mV.

At 500 kN the expected output is 10.0 mV. The instrument reads 9.8 mV — the error is mV, that is .

If the calibration certificate gives a linearity of ±0.25 % of full scale, the reading is out of tolerance: the cell, cable or amplifier is checked and re-calibrated before use.

Calibration interval and records

Instruments are recalibrated at regular intervals (typically once a year, or before and after a major test), after any shock, overload or repair, and whenever a result looks doubtful. Each sensor has a calibration record (serial number, date, factor, uncertainty, the next due date), which is kept with the channel map of the system.

This chapter is in the syllabus of

Open an exam to see where this chapter sits in its syllabus, and to practise it.